Electrode Design
SIDE VIEW OF S650CD ANGLED VIEW OF BOTTOM/
SIDE OF S650CD
BOTTOM VIEW OF S650CD


Electrode Mounting Choices
SIDE STREAM(BY-Pass) LINE MOUNTING
  Flow cell's 1/4" high flow channel provides 2 ft/sec velocity for each GPM of flow. 3 to 5 GPM flows recommended for most applications.

INSERTION MOUNTING
  Assemblies are available with 12, 18 and 24 inch insertion depths.

SUBMERSION MOUNTING IN-LINE MOUNTING



Temperature Compensation

pH TEMPERATURE ERROR TABLE
¡É
pH
2
3
4
5
6
7
8
9
10
11
12
5
.30
.24
.18
.12
.06
0
.06
.12
.18
.24
.30
15
.15
.12
.09
.06
.03
0
.03
.06
.09
.12
.15
25
0
0
0
0
0
0
0
0
0
0
0
35
.15
.12
.09
.06
.03
0
.03
.06
.09
.12
.15
45
.30
.24
.18
.12
.06
0
.06
.12
.18
.24
.30
55
.45
.36
.27
.18
.09
0
.09
.18
.27
.36
.45
65
.60
.48
.36
.24
.12
0
.12
.24
.36
.48
.60
75
.75
.60
.45
.30
.15
0
.15
.30
.45
.60
.75
85
.90
.72
.54
.36
.18
0
.18
.36
.54
.72
.90

     0 pH Error Range       Less than .1 pH Error Range


SYSTEM SPECIFICATIONS

pH RANGE:
  0-14 pH (0-12 without sodium ion error.)
SPEED OF RESPONSE:
  95% in less than 5 seconds.
TEMPERATURE/PRESSURE TATING:

(C¢ª) (F¢ª) (PSIG)
10-76 50-170 100
10-81 50-180 85
10-100 50-212 50

MATERIALS WETTED BY SAMPE:

CPVC DESIGNS: CPVC ployethylene, viton and glass (and platinum for ORP electrodes).

PVDF DESIGNS: PVDF (kyanr), viton and glass (and platinum for ORP electrodes).




Ordering Information
SIDE STREAM ORDER 3 ITEMS DESCRIPTION CPVC PARTS PVDF PARTS
  Systems
  without
  Temperature
  Compensation
1. Flow Cell
2. Select one
    Electrode
3. Select one
    Cable Assembly
Flow Cell................................................................
pH Combination Electrode.........................................
ORP (REDOX) Combination Electrode.........................
For conduit use........................................................
For exposed cable use.............................................
*SPECIFY needed cable length and connector.
FC60C
S645CD
S645CD-ORP
S653*
S648*
FC60K
S645KD
S645KD-ORP
Use CPVC Part
Use CPVC Part
SIDE STREAM ORDER 2 ITEMS DESCRIPTION CPVC PARTS PVDF PARTS
  Systems
  with
  Temperature
  Compensation
1. Electrode
2. Flow Cell
   with cable
   Assembly
pH Combination Electrode.........................................
Flow Cell and S648 Cable Assembly
modified to include Temperature Compensator............
*SPECIFY Temperature Compensator Code (See Automatic Temperature Compensator Selection Chart), needed cable length and connector.
S645CD

FC60TC-£ß*
S645KD

FC60TK-£ß*
SUBMERSION ORDER 2 ITEMS DESCRIPTION CPVC PARTS PVDF PARTS
  Systems
  without
  Temperature
  Compensation
1. Select one
    Electrode
2. Cable Assembly
pH Combination Electrode.........................................
ORP (REDOX) Combination Electrode.........................
For Submersion use.................................................
*SPECIFY needed cable length and connector.
S650CD
S650CD-ORP
S653*
S650KD
S650KD-ORP
S653K*
SUBMERSION ORDER 2 ITEMS DESCRIPTION CPVC PARTS PVDF PARTS
  Systems
  with
  Temperature
  Compensation
1. Electrode
2. Cable Assembly
pH Combination Electrode.........................................
S653 Cable Assembly modified to
include Temperature Compensator.............................
*SPECIFY Temperature Compensator Code (See Automatic Temperature Compensator Selection Chart), needed cable length and connector.
S650CD

S653TC-£ß*
S650KD

S653TK-£ß*
INSERTION ORDER 3 ITEMS DESCRIPTION CPVC PARTS PVDF PARTS
  Systems
  without
  Temperature
  Compensation
1. Select one
   Insertion
   Assembly
2. Select one
    Electrode
3. Select one
    Cable Assembly
Insertion Assembly, 12" insertion depth.......................
Insertion Assembly, 18" insertion depth.......................
Insertion Assembly, 24" insertion depth.......................
pH Combination Electrode.........................................
ORP (REDOX) Combination Electrode.........................
For conduit use........................................................
For exposed cable use.............................................
*SPECIFY needed cable length and connector.
S675
S676
S677
S655CD
S655CD-ORP
S653*
S648*
S675K
Not available
Not available
S655KD
S655KD-ORP
Use CPVC Part
Use CPVC Part
INSERTION ORDER 2 ITEMS DESCRIPTION CPVC PARTS PVDF PARTS
  Systems
  with
  Temperature
  Compensation
1. Electrode
2. Insertion Assembly
pH Combination Electrode.........................................
Insertion Assembly, 12" insertion depth
with cable & temperature comoensator.......................
*SPECIFY Temperature Compensator Code (See Automatic Temperature Compensator Selection Chart), needed cable length and connector.
S655CD

S675TC-£ß*
S655KD

S675TK-£ß*
IN-LINE ORDER 3 ITEMS DESCRIPTION CPVC PARTS  
Systems
  without
  Temperature
  Compensation
1. Select one
    Electrode
2. Gland
3. Select one
    Cable Assembly
pH Combination Electrode.........................................
REDOX Combination Electrode...................................
3/4" Male NPT Electrode Gland..................................
For conduit use........................................................
For exposed cable use.............................................
*SPECIFY needed cable length and connector.
S660CD
S660CD-ORP
FC66P
S653*
S648*
 
IN-LINE ORDER 2 ITEMS DESCRIPTION CPVC PARTS  
  Systems
  with
  Temperature
  Compensation
1. Electrode
2. Gland
pH Combination Electrode.........................................
Insertion Assembly, 12" insertion depth
with cable & temperature comoensator.......................
*SPECIFY Temperature Compensator Code (See Automatic Temperature Compensator Selection Chart), needed cable length and connector.
S660CD

FC66TC£ß*